MR-A2600 Hospital Anesthesia Machine
Special Feature
1. Patient type: Adult, Pediatric
2. Driven mode: Pneumatically driven electronically controlled
3. Mode: Semi-open, Semi-closed or Closed system
4. Aesthetic vaporizer: one vaporizer
5. Power supply:220v±10%, 50Hz±1Hz
110v±10%, 60Hz±1Hz
6. Power consumption:≤50W
7. Internal battery: DC12V (over 4 working hours)
Anesthesia ventilator
1. Respiration mode: IPPV, SIMV, SIGH, MANUAL
2. Display: 5.5' SNT screen
3. Waveform: P-t, V-t
4. Tidal volume: 50ml~1600ml
5. Respiratory frequency: 1~65BPM
6. I:E ratio: 8:1~1:8
7. Alarm parameters: High pressure alarm, Low pressure alarm, O2 source failure alarm
8. Pressure alarm setting:
High pressure alarm limit: 20~60cmH2O
Low pressure alarm limit: 0~20cmH2O
9. Monitor parameters: Tidal volume, MV, Ppeak, PEEP, respiratory frequency, I:E ratio
10. Safety valve pressure limit: 6 Kpa
11. Good air tightness, high reliability, and minor respiration impedance
Main unit
1. Circle absorber: compact circle absorber
2. Bellow: 0~1600ml
3. Flow meter:
O2, N2O four-tube flow meter
O2 tube: 0L/min~1L/min~10L/min
N2O tube: 0L/min~1L/min~10L/min
O2 and N2O Linkage Device Adjust O2 and N2O in proportion(keep O2 concentration not lower than 25%)
4. Oxygen FLUSH: 25L/min~75L/min
5. Gas source pressure:
O2: 0.25Mpa~0.65Mpa
N2O: 0.25Mpa~0.65Mpa
6. Gauge: Two gauges, N2O, O2